Title:
De helixtest in de praktijk
Authors:
Bruijn de ACP; Drongelen van AW
Other Titles:
Application of the helix test in hospital sterilisers
Abstract:
The RIVM performed the helix test in the autoclaves of 20 hospitals. Only 59% of the test results were positive, while the Bowie and Dick test was positive in all cases. The test helix added to a light instrument tray proved to be a particular challenge to the sterilisation process. One of the measures for infection prevention in hospital is the sterilisation of the surgical instruments by steam sterilisation. For this sterilisation method it is important that steam is able to get to all surfaces of the instruments and to penetrate into hollow instruments. European standardisation committee CEN TC102 developed the helix test to demonstrate steam penetration into hollow devices. This method utilises a long narrow tube. Steam penetration throughout the length of the tube is demonstrated with an indicator at the end of the tube. This test method is essentially different from the currently used test method that utilises a textile pack; the Bowie and Dick test. Nowadays the vast majority of the hospitals do not sterilise textile packs, but there is a distinct increase in the number of hollow instruments that is reprocessed, eg. laparoscopic instruments. Hospitals should adopt their sterilisation indicators to the type of steriliser load and are advised to evaluate the performance of the sterilisers with the helix test added to a light instrument tray.

Om na te gaan in hoeverre de autoclaven op de centrale sterilisatie afdelingen in de Nederlandse ziekenhuizen in staat zijn om de helixtest conform Europese norm EN867-5 tot een positief resultaat te brengen is een praktijkonderzoek uitgevoerd in 20 ziekenhuizen. In totaal zijn 476 testresultaten verkregen, waarvan slechts 59% positief was. Na analyse van de testresultaten blijkt de uitslag van de testhelixen bepaald te worden door het type testhelix, het type lading waarin de testhelix wordt toegevoegd en het type sterilisatieproces. Twee van de vier typen testhelixen geven minder vaak een positief testresultaat. Alle typen testhelixen komen minder vaak tot een positief resultaat indien ze worden opgenomen in een testlading. Een licht instrumentennet met een massa van 2 kg bleek in dit kader het meest kritisch te zijn. Het type sterilisatieproces waarin gebruikt wordt gemaakt van de 'klassieke' ontluchtingsfase, bestaande uit drie diepe vacuumpulsen afgewisseld met stoompulsen, blijkt een significant beter resultaat te geven dan andere type sterilisatieprocessen. De helixtest kan een waardevolle bijdrage leveren aan de optimalisatie van het sterilisatieproces en de stoomkwaliteit ten behoeve van de sterilisatie van holle instrumenten. Het rapport geeft informatie en aanbevelingen om tot implementatie van de helixtest te komen.
Affiliation:
BMT
Publisher:
Rijksinstituut voor Volksgezondheid en Milieu RIVM
Issue Date:
6-Aug-2004
URI:
http://hdl.handle.net/10029/8848
Additional Links:
http://www.rivm.nl/bibliotheek/rapporten/265011001.html
Language:
nl
Series/Report no.:
RIVM Rapport 265011001
Appears in Collections:
RIVM reports - old archive

Full metadata record

DC FieldValue Language
dc.contributor.authorBruijn de ACPen_US
dc.contributor.authorDrongelen van AWen_US
dc.date.accessioned2007-02-23T15:28:05Z-
dc.date.available2007-02-23T15:28:05Z-
dc.date.issued2004-08-06en_US
dc.identifier265011001en_US
dc.identifier.urihttp://hdl.handle.net/10029/8848-
dc.description.abstractThe RIVM performed the helix test in the autoclaves of 20 hospitals. Only 59% of the test results were positive, while the Bowie and Dick test was positive in all cases. The test helix added to a light instrument tray proved to be a particular challenge to the sterilisation process. One of the measures for infection prevention in hospital is the sterilisation of the surgical instruments by steam sterilisation. For this sterilisation method it is important that steam is able to get to all surfaces of the instruments and to penetrate into hollow instruments. European standardisation committee CEN TC102 developed the helix test to demonstrate steam penetration into hollow devices. This method utilises a long narrow tube. Steam penetration throughout the length of the tube is demonstrated with an indicator at the end of the tube. This test method is essentially different from the currently used test method that utilises a textile pack; the Bowie and Dick test. Nowadays the vast majority of the hospitals do not sterilise textile packs, but there is a distinct increase in the number of hollow instruments that is reprocessed, eg. laparoscopic instruments. Hospitals should adopt their sterilisation indicators to the type of steriliser load and are advised to evaluate the performance of the sterilisers with the helix test added to a light instrument tray.en
dc.description.abstractOm na te gaan in hoeverre de autoclaven op de centrale sterilisatie afdelingen in de Nederlandse ziekenhuizen in staat zijn om de helixtest conform Europese norm EN867-5 tot een positief resultaat te brengen is een praktijkonderzoek uitgevoerd in 20 ziekenhuizen. In totaal zijn 476 testresultaten verkregen, waarvan slechts 59% positief was. Na analyse van de testresultaten blijkt de uitslag van de testhelixen bepaald te worden door het type testhelix, het type lading waarin de testhelix wordt toegevoegd en het type sterilisatieproces. Twee van de vier typen testhelixen geven minder vaak een positief testresultaat. Alle typen testhelixen komen minder vaak tot een positief resultaat indien ze worden opgenomen in een testlading. Een licht instrumentennet met een massa van 2 kg bleek in dit kader het meest kritisch te zijn. Het type sterilisatieproces waarin gebruikt wordt gemaakt van de 'klassieke' ontluchtingsfase, bestaande uit drie diepe vacuumpulsen afgewisseld met stoompulsen, blijkt een significant beter resultaat te geven dan andere type sterilisatieprocessen. De helixtest kan een waardevolle bijdrage leveren aan de optimalisatie van het sterilisatieproces en de stoomkwaliteit ten behoeve van de sterilisatie van holle instrumenten. Het rapport geeft informatie en aanbevelingen om tot implementatie van de helixtest te komen.nl
dc.format.extent423000 bytesen_US
dc.format.extent432183 bytes-
dc.format.mimetypeapplication/pdf-
dc.language.isonlen_US
dc.publisherRijksinstituut voor Volksgezondheid en Milieu RIVMen_US
dc.relation.ispartofseriesRIVM Rapport 265011001en_US
dc.relation.urlhttp://www.rivm.nl/bibliotheek/rapporten/265011001.htmlen_US
dc.subject.othersterilisationen
dc.subject.othersteamen
dc.subject.othersteam penetrationen
dc.subject.otherhelix testen
dc.subject.otherhollow loaden
dc.subject.otherprocess challenge deviceen
dc.subject.otherbowie & dick testen
dc.subject.otherchemical indicatoren
dc.subject.othersterilisatienl
dc.subject.otherstoomnl
dc.subject.otherstoompenetratienl
dc.subject.otherhelixtestnl
dc.subject.otherholle instrumentennl
dc.subject.otherbowie & dick testnl
dc.subject.otherchemische indicatornl
dc.titleDe helixtest in de praktijken_US
dc.title.alternativeApplication of the helix test in hospital sterilisersen_US
dc.contributor.departmentBMTen_US
All Items in WARP are protected by copyright, with all rights reserved, unless otherwise indicated.