Bepaling van de dosis(tempo) - verdeling in de biologische bestralingsfaciliteit van het RIVM
Citations
Altmetric:
Series / Report no.
Open Access
Type
Report
Language
nl
Date
1990-02-28
Research Projects
Organizational Units
Journal Issue
Title
Bepaling van de dosis(tempo) - verdeling in de
biologische bestralingsfaciliteit van het RIVM
Translated Title
Determination of the absorbed dose and dose
distribution in the biological irradiation facility of the
RIVM
Published in
Abstract
Abstract niet beschikbaar
The RIVM has available a biological irradiation facility (BBF) for the irradiation of biological samples and small animals. The dosimetry of the BBF was performed by the Dosimetry Section of the RIVM. The results show that the absorbed dose in water (determined on January 1, 1990) is equal to 0.97 gy/timer-unit, with a total uncertainty of 7% (1 sigma). During the rotation of the irradiation chamber the absorbed dose (determined on January 1, 1990) is equal to 0.38 gy, with a total uncertainty of 15% (1 sigma).
The RIVM has available a biological irradiation facility (BBF) for the irradiation of biological samples and small animals. The dosimetry of the BBF was performed by the Dosimetry Section of the RIVM. The results show that the absorbed dose in water (determined on January 1, 1990) is equal to 0.97 gy/timer-unit, with a total uncertainty of 7% (1 sigma). During the rotation of the irradiation chamber the absorbed dose (determined on January 1, 1990) is equal to 0.38 gy, with a total uncertainty of 15% (1 sigma).
Description
Publisher
Sponsors
RIVM