Prestatiekenmerken van de bepaling van As, Cd, Pb en Hg in voeding en biologisch materiaal met ICP-MS en AFS
Velde-Koerts T van der ; Bom CM
Velde-Koerts T van der
Bom CM
Citations
Altmetric:
Series / Report no.
Open Access
Type
Report
Language
nl
Date of publication
1999-03-01
Year of publication
Research Projects
Organizational Units
Journal Issue
Title
Prestatiekenmerken van de bepaling van As, Cd, Pb en
Hg in voeding en biologisch materiaal met ICP-MS en
AFS
Translated Title
Performance characteristics for the determination
of As, Cd, Hg and Pb in food and biological using ICP-MS and
AFS
Published in
Abstract
Een ICP-MS-methode werd ontwikkeld voor de bepaling van
As, Cd en Pb in salpeterzuur-destruaten van voeding en biologisch materiaal.
De methode werd geoptimaliseerd op basis van minimale spectrale storingen en
matrixeffecten met behulp van correctieformules, interne standaarden en
matrixmatching. De bepaling van Hg werd uitgevoerd met een AFS-methode. De
volgende aantoonbaarheidsgrenzen werden gevonden: 0.01 microg/g As, 0.005
microg/g Cd, 0.02 microg/g Pb en 0.0007 microg/g Hg. De relatieve
herhaalbaarheidsstandaarddeviatie was beter dan 7%. Omdat de terugvinding
voor As systematisch hoger was dan 110% (tot 133%) werden de concentraties
voor As, Cd en Pb gecorrigeerd voor de terugvinding van een geaddeerde
standaard. De juistheid van referentie-materialen lag tussen 87-106% voor
As, Cd, Pb en Hg. De prestatiekenmerken van beide methoden waren goed en
ICP-MS en AFS zijn derhalve geschikt voor de kwantitatieve analyse van
salpeterzuurdestruaten van voeding en biologisch
materiaal.
The ICP-MS method developed for the determination of As, Cd, Pb in nitric acid digests of food and biological material was optimized for minimal spectral interferences and matrix effects using interference corrections, internal standards and matrix matching of the calibration standards. In the AFS method used for the determination of Hg, the following detection limits were found: 0.01 g/g As, 0.005 g/g Cd, 0.02 g/g Pb and 0.0007 g/g Hg. The repeatability was better than 7% RSD. Since the recovery for As was systematically too high (110% to 133%), concentrations were corrected for recovery. The trueness of certified reference materials ranged from 87% to 106%. In both methods performance was good, making them suitable for the quantitative analysis of nitric acid digests of food and biological materialis of nitric acid digests of food and biological material.
The ICP-MS method developed for the determination of As, Cd, Pb in nitric acid digests of food and biological material was optimized for minimal spectral interferences and matrix effects using interference corrections, internal standards and matrix matching of the calibration standards. In the AFS method used for the determination of Hg, the following detection limits were found: 0.01 g/g As, 0.005 g/g Cd, 0.02 g/g Pb and 0.0007 g/g Hg. The repeatability was better than 7% RSD. Since the recovery for As was systematically too high (110% to 133%), concentrations were corrected for recovery. The trueness of certified reference materials ranged from 87% to 106%. In both methods performance was good, making them suitable for the quantitative analysis of nitric acid digests of food and biological materialis of nitric acid digests of food and biological material.