    Ionchromatografische analyse van enkele organische anionen

    • CSV
    • RefMan
    • EndNote
    • BibTex
    • RefWorks
    Average rating
     
       votes
    Cast your vote
    You can rate an item by clicking the amount of stars they wish to award to this item. When enough users have cast their vote on this item, the average rating will also be shown.
    Star rating
     
    Your vote was cast
    Thank you for your feedback
    Authors
    Neele J
    Cleven RFMJ
    Jonker N
    Versteegh JFM
    Type
    Report
    Language
    nl
    
    Metadata
    Show full item record
    Title
    Ionchromatografische analyse van enkele organische anionen
    Translated Title
    Ion chromatographic analysis of some organic anions
    Publiekssamenvatting
    Er wordt een beschrijving gegeven van de ontwikkeling, de prestatiekenmerken en de toepassing van een ionchromatografische methode voor de bepalingen van enkele organische anionen in water. Acetaat, formiaat en oxalaat worden van o.a. anorganische anionen gescheiden met een AS10 HC kolom met gradient elutie met loog. Na chemische suppressie wordt gedetecteerd met geleidbaarheids- en UV-detectie bij 192 nm. De aantoonbaarheidsgrenzen voor acetaat, formiaat en oxalaat in ruw water bedragen 0,1 umol/l. De terugvinding van de methode is getoetst in drinkwater en watertypen (grond- en oppervlaktewater) waaruit drinkwater geproduceerd wordt. De gemiddelde waarden bedragen 83, 95 en 91 % (n=24) respectievelijk voor acetaat, formiaat en oxalaat. De herhaalbaarheid voor een 1 umol/l mengstandaard van acetaat, formiaat en oxalaat bedraagt respectievelijk 5, 3 en 2% en voor een 4 umol/l mengstandaard 3, 1 en 3%. De ontwikkelde methode is toegepast in het onderzoek naar organische anionen in onder andere geozoniseerd drinkwater en watertypen waaruit drinkwater geproduceerd wordt. Voorbehandeling is voor deze analyses niet noodzakelijk gebleken. Bepalingen van ionchromatografische analyses van acetaat, formiaat en oxalaat in de diverse matrices zijn verricht. De aanwezigheid van de hoeveelheden aan monsters toegevoegd acetaat, formiaat en oxalaat kon met de methode worden bevestigd. De acetaat-, formiaat- en oxalaat- concentraties in de monsters varieerden van 0,1 tot 5,6 umol/l. De ontwikkelde methode is beschreven in een bruikbaar formaat dat in dit rapport is opgenomen. Het onderzoek werd verricht in de periode januari 2000 tot april 2000.<br>
    This rapport presents the development, the performance characteristics and the application of an ion chromatographic method for the determination of some organic anions in water. Separation of acetate, formate and oxalate is carried out with an AS10HC column and gradient elution with sodium hydroxide. After chemical suppression the components are detected with conductivty and UV-detection (192 nm). The performance characteristics are determined are the precision, detections limits, resolutions and recovery data. The repeatability for acetate, formate and oxalate are is 5, 3 and 2 % for a mixed standard of 1 umol/l and 3, 1 and 3% for 4 umol/l respectively. The detection limits for acetate, formate and oxalate are 0,1umol/l 6, 5 and 9 ug/l respectively. Results for recovery experiments are adequate with average values of 83, 95 en 92 % for acetate formate and oxalate respectively (n=24). The method was applied in the studies of small organic anions in samples of production of drinking water treated with ozonated drinking waters. Small amounts, till 0,20 mg/l (3,3 umol/l) acetate 0,25 mg/l (5,6 umol/l) formate 0,25 mg/l (2,8 umol/l) and oxalate are build by ozonisation of water. The resulting analytical procedure of this study is described in a usefull format.<br>
    Publisher
    Rijksinstituut voor Volksgezondheid en Milieu RIVM
    Sponsors
    RIVM
    Collections
    RIVM official reports

    entitlement

     

    DSpace software (copyright © 2002 - 2023)  DuraSpace
    Quick Guide | Contact Us
    Open Repository is a service operated by 
    Atmire NV
     

    Export search results

    The export option will allow you to export the current search results of the entered query to a file. Different formats are available for download. To export the items, click on the button corresponding with the preferred download format.

    By default, clicking on the export buttons will result in a download of the allowed maximum amount of items.

    To select a subset of the search results, click "Selective Export" button and make a selection of the items you want to export. The amount of items that can be exported at once is similarly restricted as the full export.

    After making a selection, click one of the export format buttons. The amount of items that will be exported is indicated in the bubble next to export format.