Geautomatiseerde monsterzuivering door kolomschakeling bij de bepaling van lorazepam in ratteserum met hoge-prestatie vloeistofchromatografie
Average rating
Cast your vote
You can rate an item by clicking the amount of stars they wish to award to this item.
When enough users have cast their vote on this item, the average rating will also be shown.
Star rating
Your vote was cast
Thank you for your feedback
Thank you for your feedback
Type
ReportLanguage
nl
Metadata
Show full item recordTitle
Geautomatiseerde monsterzuivering door kolomschakeling bij de bepaling van lorazepam in ratteserum met hoge-prestatie vloeistofchromatografieTranslated Title
Automated sample clean-up by column switching in the determination of lorazepam in rat serum with high-performance liquid chromatographyPubliekssamenvatting
Abstract niet beschikbaarIn this report a study is described on the application of automated sample clean up by column switching in the quantitative determination of lorazepam in rat serum. Extraction of lorazepam from serum is by using a solid-phase extraction procedure with an "High-Hydrophobic" octadecylsilane sorbent. Determination of lorazepam is with high- performance liquid chromatography. The solid-phase extract is separated over a precolumn from serum components by "on line" column switching. Detection is by UV-detection at 230 nm. The extraction recovery (+- standard deviation) is (91 +- 0.7)%. The precision is 2.4% at 1.12 mug/l. The method is adapted to a concentration range from 0.05 to 4.2 mug/l.
Sponsors
HIG CBGCollections