    Kwaliteitscontrole interim meetnet ammoniak: een onderzoek naar de invloed van de temperatuur in de meetstations

    • CSV
    • RefMan
    • EndNote
    • BibTex
    • RefWorks
    Average rating
     
       votes
    Cast your vote
    You can rate an item by clicking the amount of stars they wish to award to this item. When enough users have cast their vote on this item, the average rating will also be shown.
    Star rating
     
    Your vote was cast
    Thank you for your feedback
    Authors
    Elzakker BG van
    Stuiver J
    Uden GJBM van
    Type
    Report
    Language
    nl
    
    Metadata
    Show full item record
    Title
    Kwaliteitscontrole interim meetnet ammoniak: een onderzoek naar de invloed van de temperatuur in de meetstations
    Translated Title
    Quality Control in the interim network for the measurement of ammonia: an investigation of the influence of the temperature inside the measurement stations
    Publiekssamenvatting
    I.v.m. aanwijzingen dat de temperatuurscorrectie voor de meetapparatuur zoals in gebruik in het interim meetnet ammoniak niet correct zou zijn is (opnieuw) onderzoek uitgevoerd naar de temperatuurgevoeligheid van deze apparaten met zowel de 'oude' als de 'nieuwe', door de fabrikant aangegeven correctie. De 'oude' correctiemethode is fout daar deze uitgaat van de verkeerde referentietemperatuur terwijl de 'nieuwe' correctiemethode wel uitgaat van de goede referentietemperatuur (20 graden C). Daarnaast is onderzocht wat aanpassing (volgens de 'nieuwe' correctiemethode) van het bestaande bestand aan ammoniakdata uit het interim meetnet voor gevolgen zou hebben voor individuele uurwaarden, periodegemiddelde concentraties en het periodegemiddelde dagverloop.Tevens is inzicht verkregen in de werkelijke temperatuurrange in de geconditioneerde meetstations. Het klimaatkameronderzoek naar de temperatuurgevoeligheid toont afwijkingen in de 1-uursgemiddelde concentraties tot -25% bij 35 graden C (26 mug/m3, 'oude', huidige correctiemethode) resp. +20% bij 35 graden C (1.3 mug/m3, 'nieuwe' correctiemethode). Het onderzoek toont met name aan dat rekening is te houden met een concentratieafhankelijkheid van de temperatuurgevoeligheid. Gebeurd dit niet (zoals bij de 'nieuwe' correctiemethode) dan kunnen bij lage concentraties (ca. 1 mug/m3) nog significante afwijkingen optreden van -11% tot +20% bij temperaturen van 10 resp. 35 graden C ('nieuwe' correctiemethode). Onder dit concentratieniveau wordt de absolute afwijking echter marginaal. De temperatuur in de meetbehuizingen is in de winter 20 +-2 graden C. In de zomer bij hoge buitenluchttemperaturen (30 graden C) blijkt de temperatuur in de meethut rond 19.00 uur op te kunnen lopen tot zo'n 35 graden C. De airconditioning is dus ondergedimensioneerd. Met de huidige 'oude' temperatuurscorrectie worden bij deze hoge temperaturen uurgemiddelde ammoniakconcentraties zo'n 25 % onderschat. Voor periodegemiddelde (2 tot 4 weken) concentraties blijft deze onderschatting beperkt tot ca.3 %. Dit wordt veroorzaakt door de relatief lage frequentie van gelijktijdig voorkomen van hoge temperaturen en hoge concentraties. Om diezelfde reden wordt het periodegemiddelde dagprofiel van NH3 ook slechts marginaal beinvloed. Aanbevolen wordt (na aanvullend onderzoek en indien praktisch mogelijk) de temperatuurscorrectie concentratie afhankelijk te maken en/of de temperatuurshuishouding in de meetbehuizingen beter te beheersen en/of thermostatisering van de apparatuur.
    Because of doubts concerning the correction method for temperature sensitivitiy of the instruments used in the interim network for the measurement of ammonia (NH3), this sensitivity was investigated again using two correction methods ("old" and "new") as proposed by the manufacturer. The "old" method is used in the interim network and turned out to be principally wrong because it does not use the correct reference temperature while the "new" method does, namely 20 degrees C. Also, the influence on 1-hour averages, period-averages and diurnal variations when using the "new" method for recalculating the excisting NH3 database, which was determined with the "old" correction method, was investigated. A deeper understanding was gathered about the real temperature range in the measurement cabins. The investigation in a climate chamber showed deviations in the 1-hour averaged concentrations up to -25% at 35 degrees C (26 mug/m3, "old" correction method) and +20% at 35 degrees C (1.3 mug/m3, "new" correction method) respectivily. Particularly, a concentration dependence was found for the temperature sensitivity. If no concentration depending correction is used, deviations of -11% up to +20% at temperatures of 10 and 35 degrees C respectively ("new" correction method) can still occur particularly with low concentrations (ca. 1 mug/m3). Below this value, the absolute deviation becomes small. During wintertime, the temperature in the cabins is 20 +-2 degrees C. In summer, at high ambient temperatures (30 degrees C), the temperature in the cabin can reach 35 degrees C at about 19.00 h. So, the air-conditioning is not dimensioned well. With the "old" correction method, which is now in use, one hour averaged ammonia concentrations are underestimated with about 25% at these high temperatures. Period averages (2 - 4 weeks) are underestimated less, namely about 3%. This is caused by the low frequency of simultaneous occurrence of high temperatures and high concentrations. For the same reason, diurnal variations are only marginally influenced. It is recommended (after additional investigation) to make the temperature correction concentration dependent, to improve the temperature-conditioning in the cabins or to thermostat future versions of the instrument itself.
    Sponsors
    RIVM (DGM/LE
    URI
    http://hdl.handle.net/10029/261934
    Collections
    RIVM official reports

    entitlement

     

    DSpace software (copyright © 2002 - 2023)  DuraSpace
    Quick Guide | Contact Us
    Open Repository is a service operated by 
    Atmire NV
     

    Export search results

    The export option will allow you to export the current search results of the entered query to a file. Different formats are available for download. To export the items, click on the button corresponding with the preferred download format.

    By default, clicking on the export buttons will result in a download of the allowed maximum amount of items.

    To select a subset of the search results, click "Selective Export" button and make a selection of the items you want to export. The amount of items that can be exported at once is similarly restricted as the full export.

    After making a selection, click one of the export format buttons. The amount of items that will be exported is indicated in the bubble next to export format.