    Onderzoek naar 11 ammoniak monitoren voor het interim meetnet ammoniak

    • CSV
    • RefMan
    • EndNote
    • BibTex
    • RefWorks
    Average rating
     
       votes
    Cast your vote
    You can rate an item by clicking the amount of stars they wish to award to this item. When enough users have cast their vote on this item, the average rating will also be shown.
    Star rating
     
    Your vote was cast
    Thank you for your feedback
    Authors
    Elzakker BG van
    Stuiver J
    Uden GJBM van
    Uiterwijk JW
    Type
    Report
    Language
    nl
    
    Metadata
    Show full item record
    Title
    Onderzoek naar 11 ammoniak monitoren voor het interim meetnet ammoniak
    Translated Title
    Investigation of 11 instruments for the measurement of ammonia in the interim network
    Publiekssamenvatting
    Onderzocht zijn een 11-tal monitoren voor ammoniak (NH3)- immissiemetingen in het Landelijk Meetnet Luchtkwaliteit (LML). Deze continue-flow natte denuders (type AMANDA) zijn ontwikkeld en gefabriceerd door het Energieonderzoek Centrum Nederland (ECN) te Petten. De apparatuur is onder laboratorium condities onderzocht op lineariteit, afwijking, herhaalbaarheid, detectiegrens, precisie, gevoeligheid voor interferenten, temperatuurgevoeligheid, vochtgevoeligheid en gevoeligheid voor netspanningsvariaties. De resultaten zijn als volgt samen te vatten: -de lineariteit in het meetbereik 0 tot 250 mug/m3 is zeer goed (r=1.000) -afwijkingen t.o.v. aangeboden ammoniakconcentraties liggen gemiddeld voor alle instrumenten tussen maximaal +3.5% voor concentraties < 30 mug/m3 NH3 en maximaal -7.3% voor NH3 concentraties > 30 mug/m3. Incidenteel kunnen grotere afwijkingen voorkomen. -de herhaalbaarheid bij nullucht ligt in de orde van 0.5 mug/m3 en bij een concentratie van 168 mug/m3 in de orde van 2% -de detectiegrens (2s) bedraagt 0.05 +- 0.04 mug/m3 -de precisie bij nullucht is 0.03 +-0.02 mug/m3 en voor NH3 concentraties < 2% -de temperatuurgevoeligheid is significant in het onderzochte temperatuurstraject van 10 tot 30 graden C -er is geen interferentie vast te stellen voor concentraties van 86.5 mug/m3 SO2, 135 mug/m3 NO, 207 mug/m3 NO2 en 120 mug/m3 O3 -een geringe vochtgevoeligheid is vastgesteld maar wordt waarschijnlijk geintroduceerd door artefacten buiten de apparatuur -de apparatuur vertoont geen significante gevoeligheid voor te verwachten netspanningsvariaties (200 - 240V) Aanbevolen wordt de apparatuur uitsluitend onder zeer strikte condities in het LML te bedrijven met inzet van veel personele capaciteit van voldoende deskundigheidsniveau. De opzet van een kwaliteitscontroleprogramma is noodzakelijk om de kwaliteit onder veldcondities vast te stellen. Daarnaast wordt aanbevolen aanvullend onderzoek naar de temperatuursgevoeligheid van de apparatuur uit te voeren.
    Eleven instruments for ammonia (NH3) ambient concentration measurements have been investigated under laboratory conditions for temporary use in the National Air Quality Monitoring Network (LML). These so-called continuous-flow wet denuders (type AMANDA) have been developed and manufactured by the Netherlands Energy Research Foundation (ECN). The results of the investigation can be summarized as follows: -the linearity in the measurement range from 0 to 250 mug/m3 NH3 is excellent (r=1.000) -deviations differ (all instruments average) from +3.5% (concentrations < 30 mug/m3 NH3) to -7.3% (concentrations > 30 mug/m3) ; greater deviations may occur occasionally. -repeatability with zero air is about 0.5 mug/m3 and with a concentration of 168 mug/m3 about 2% - the detection limit (2s) is 0.05 +- 0.04 mug/m3 -precision at zero air is 0.03 +-0.02 mug/m3 and for NH3 concentrations it is < 2% - temperaturesensitivity is significant in the investigated temperature range of 10 to 30 degrees C -no interference is determined for concentrations of 86.5 mug/m3 SO2, 135 mug/m3 NO, 207 mug/m3 NO2 and 120 mug/m3 O3 -a small sensitivity for humid air is determined ; however this is probably introduced by artefacts outside the instruments -the instruments show no significant sensitivity for line-voltage variations between 200 and 240V It is recommended to use these instruments only under strict conditions in LML with sufficient personal capacity. The development of a quality control program will be necessary to guarantee the quality under field conditions. Furthermore it is recommended to investigate the temperature sensitivity of the instruments.
    Sponsors
    RIVM (DGM/LE)
    URI
    http://hdl.handle.net/10029/260947
    Collections
    RIVM official reports

    entitlement

     

    DSpace software (copyright © 2002 - 2023)  DuraSpace
    Quick Guide | Contact Us
    Open Repository is a service operated by 
    Atmire NV
     

    Export search results

    The export option will allow you to export the current search results of the entered query to a file. Different formats are available for download. To export the items, click on the button corresponding with the preferred download format.

    By default, clicking on the export buttons will result in a download of the allowed maximum amount of items.

    To select a subset of the search results, click "Selective Export" button and make a selection of the items you want to export. The amount of items that can be exported at once is similarly restricted as the full export.

    After making a selection, click one of the export format buttons. The amount of items that will be exported is indicated in the bubble next to export format.