Analytisch-chemische aspekten van de aluminiumbepaling in grond- en drinkwater met ICP-MS
Citations
Series / Report no.
Open Access
Type
Language
Date
Research Projects
Organizational Units
Journal Issue
Title
Translated Title
Published in
Abstract
Abstract niet beschikbaar
The aim of this project was to investigate whether ICP-MS is a suitable method for determining aluminium at low concentration levels in ground and drinking water. The determination was performed at m/z=27 with indium as an internal standard. The calibration curve was linear up to at least 100 mug/l A1 ; the detection limit (3s; n=25) was 0.4 mug/l. This detection limit could only be achieved if labware was thoroughly cleaned with 1% HNO3, if solutions were kept in polyethene bottles and if the metal parts of the autosampler were cleaned or replaced by synthetic materials. The short term stability (RSD) was 1.8% and the long term stability (RSD) was 3.0%. Spectral interferences and matrixeffects due to the most abudant ions in ground and drinking water and interferences like 11B16O+, 1H12C14N+ and 26Mg1H+ could be neglected. Only the addition of 400 mg/l sulphate resulted in an 10% positive shift in sensitivity. ICP-MS results compared well with ICP-AES results for concentrations above the detection limit of ICP-AES (20 mug/l). 70% of the 109 analyzed ground and drinking water samples (from the Netherlands) had concentrations within the linear working range (0.8-200 mug/l).