Optimalisatie van de analyse van konings-waterdestruaten met axiaal-ICP-AES
Citations
Altmetric:
Series / Report no.
Open Access
Type
Report
Language
nl
Date
1997-12-31
Research Projects
Organizational Units
Journal Issue
Title
Optimalisatie van de analyse van
konings-waterdestruaten met axiaal-ICP-AES
Translated Title
Optimalisation of the analysis of aqua regia
digests with axial ICP-AES
Published in
Abstract
Een axiaal-ICP-AES-methode werd ontwikkeld voor de
analyse van 25 elementen in konings-waterdestruaten van milieumonsters. De
methode werd geoptimaliseerd op basis van: plasma-instellingen,
meetcondities, correctieprocedures voor spectrale storingen, spoeltijden,
standaardsamenstelling, interne standaarden en kalibratieschema's. De
voorgestelde analysemethode is geschikt voor de elementanalyse van
bodemmateriaal en asresten. K, Na, S, Si en Zn kunnen alleen indicatief
bepaald worden. De aantoonbaarheidsgrenzen voor Ba, Co, Cr, Cu, Mo, Ni en
Pb in bodemmateriaal liggen op of onder het niveau dat voor het AP04-besluit
vereist is; voor As, Cd en Hg is de aantoonbaarheidsgrens niet voldoende
laag. De aantoonbaarheidsgrens voor biologisch en filtermateriaal is niet
voldoende laag, voor deze materialen dient een monster-specifieke
analysemethode ontwikkeld te worden.
An axial-ICP-AES method was developed for the analysis of 25 elements in aqua regia digests of environmental samples. The method was optimized for plasma conditions, measurement conditions, correction procedures for spectral interferences, rinse times, standard composition, internal standardization and calibration schemes. Except for K, Na, S, Si and Zn and low concentrations of As, Cd and Hg, the proposed method is suitable for elemental analysis of soil and ashes. The detection limit for biological and filter material is not sufficient; these materials can be analysed if a material specific method is going to be developed.
An axial-ICP-AES method was developed for the analysis of 25 elements in aqua regia digests of environmental samples. The method was optimized for plasma conditions, measurement conditions, correction procedures for spectral interferences, rinse times, standard composition, internal standardization and calibration schemes. Except for K, Na, S, Si and Zn and low concentrations of As, Cd and Hg, the proposed method is suitable for elemental analysis of soil and ashes. The detection limit for biological and filter material is not sufficient; these materials can be analysed if a material specific method is going to be developed.